Description
The Kowa FM-600 Laser Flare Meter provides quantitative, reproducible measurements of aqueous flare to support the clinical assessment and monitoring of ocular inflammation.
The instrument performs a non-contact, non-invasive and painless in-vivo measurement. A 635 nm diode laser scans a measurement window projected within the anterior chamber. Light scattered by proteins and other particles is detected by a photomultiplier tube and expressed numerically in photon counts per millisecond.
Unlike subjective grading methods, the FM-600 supplies objective numerical results that can help clinicians document changes and compare measurements during follow-up. It is particularly useful when evaluating uveitis, postoperative inflammation following intraocular-lens implantation and other conditions associated with increased aqueous flare.
Each measurement requires approximately 0.5 seconds. Front-view and side-view alignment modes help the operator position the measurement area accurately, while small-pupil mode supports examinations in patients with restricted pupil size.
The compact console incorporates a 5.6-inch color LCD monitor, an integrated calibrator and a 58 mm thermal printer for immediate documentation of results and graphs. Controls are positioned within the operator’s reach to facilitate a smooth workflow while facing the patient.
Key Features
- Quantitative measurement of aqueous flare
- Measurement based on laser-light scattering
- Non-contact and non-invasive examination
- Approximately 0.5-second measurement time
- Reproducible numerical results
- Measurement range of 1–500 photon counts/ms
- Front-view alignment mode
- Side-view alignment mode
- Small-pupil measurement mode
- Color alignment indicators
- Integrated calibration system
- 5.6-inch color LCD monitor
- Built-in 58 mm thermal printer
- Graphical presentation of results
- Electrically controlled vertical movement
- Patient-ID input compatibility
- RS-232C data output
- Compact ophthalmic-console design
Additional Information
| Specification | Value |
|---|---|
| Measuring laser source | Laser diode: 635 nm, 35 ± 15 μW |
| Detector | Photomultiplier tube (PMT) |
| Measuring range | 1–500 photon counts/ms |
| Measuring field | 0.3 mm vertical × 0.5 mm horizontal |
| Measuring time | 0.5 seconds |
| Working distance | 81 mm, from the examined eye to the front end of the objective lens |
| Printer | 58 mm-wide thermal line printer |
| Monitor | 5.6-inch color LCD monitor |
| Movement range | Backward/forward: 37 mm; left/right: 89 mm; vertical, electric: 25 mm |
| Dimensions | 310 (W) × 505 (D) × 462 (H) mm |
| Additional dimensions | 274 (W) × 457 (D) × 458 (H) mm |
| Weight | 22 kg / 48 lb |
| Additional weight | 18 kg |
| Interface | ID input: PS/2×, compatible with numeric keypad and barcode reader; data output: RS-232C |
| Power supply | AC 100–230 V, 50/60 Hz; 80 VA normal / 100 VA maximum |